Overview :
Filtration Probe ESIP Sterile Sampling
Bioreactor Fermenter Laboratory
Process Monitoring Production HPLC
Filtration Sampling Probe by TRACE
the solution for sterile substrate sampling


The filtration probe ESIP is designed to withdraw sterile, cell-free filtrate from fermenters and bioreactors.

The tubular micro-filtration poly-propylene membrane, supported by a stainless steel probe, is positioned inside the reaction vessel.

The filtration probe ESIP is sterilized together with the fermenter/bioreactor. Configured in either 12 mm, 19 mm or 25 mm diameters, the sampling module is installed in the vessel's standard METTLER-TOLEDO (IngoldTM) ports.

Flow rates of up to 1,5 ml/min are feasible; optimising filtration performance and minimizing sample consumption. Sterile cell-free filtrate can be withdrawn intermittently or continuously, manually or automatically from the head of the sampling device, which is outside the bioreactors or fermenters. A low-capacity peristaltic pump is typically used for withdrawal.


Features:

-  sterile Filtration; no solids > 0.2 m
-  low dead volume
-  increased safety by using a solid membrane
-  succesfully operated in various processes


Benefits:

-  no sample conditioning like centrifugation necessary
-  samples are representative
-  easy handling
-  more safety against infections by bugs
-  safety by using a solid membrane
-  to be used with on-line analyzers ProcessTRACE, MultiTRACE and ProcessMonitor
-  sampling system for on-line HPLC and other analyzers
-  can be used in small bioreactors of approx. 300 ml volume by shortening the length of the membrane from 120 mm to 40 mm. (See our news from 27/12/2025)

 

Filtration Probe
zur Fliezelle
 

magnify
picture
Technical Specifications:
 
Sampling techniquein line filtration, sterile
Membrane moduletubular PP-membrane, pre-fabricated
Adapter [mm]19, 25
Installation19, 25 mm IngoldTM standard port
Extension [mm]each 100
Flow rate [ml/min]1.5
Volume per measurement 3 ml in continuous processes