© TRACE Analytics 2008
MultiTRACE
MultiTRACE is an effective additional tool for the well proven on-line analyzing system ProcessTRACE for parallel measurement of up to six fermenters.
Always used in combination with ProcessTRACE, the MultiTRACE provides parallel measurement of the parameters glucose, methanol, ethanol, lactate or sucrose in small- volume laboratory reactor series or cubic-meter-sized production fermenters. ProcessTRACE is completely responsible for control.
Features:
- Cost-efficient and time saving optimization of medium composition and cultivation strategies
- With the industrial filtration probe or the patented dialysis probe guaranteeing sterility without sample processing
- Parallel measurement of glucose, methanol, ethanol, lactate or sucrose
Technical data:
- Measurement intervals from 5 minutes to 24 hours can be selected as desired - irrespectively of the channel
- Enzymatic-amperometric measurement principle
- Flexible measurement sequence
- Standard deviation < 3.0% for 10 g/L glucose
- Measurement range
- glucose 0.01 - 40 g/L
- lactate 0.5 - 20 g/L
- sucrose and methanol 0.5 - 40 g/L
- ethanol 1.0 - 40 g/L