Filtration probe

The filtration probe is designed to withdraw sterile, cell-free filtrate from fermenters and bioreactors.
The tubular micro-filtration polypropylene membrane, supported by a stainless steel probe, is positioned inside the reaction vessel. The filtration probe is sterilized together with the fermenter.
Configured in either 12, 19 and 25 mm diameters, the sampling module is installed in the vessels standard Mettler Toledo port. Extension modules are available for adjusting the module`s length to meet your specific requirements. The filtration unit and adapter are securely screwed together.
Sterile cell-free filtrate can be withdrawn intermittently or continuously, manually or automatically from the head of the sampling device, which is outside of the fermenter. A low-capacity peristaltic pump is typically used for withdrawal.

The filtration probe is offered in three versions:

  • Membrane length 125 mm for fermenter with a volume of 50 to 1000 Liter
  • Membrane length 40 mm for fermenter and especially glass fermenter with a volume of approximately 300 ml
  • Membrane length 125 mm with a protective sheat for fermenter with high turbulence to avoid deformation of the membrane module
Filtration probe with membrane length of 40 mm
Filtration probe with protective sheat and a membrane length of 125 mm

Features:

  • Sterile filtration; cut off 0,2 µm
  • Low dead volume
  • No sample conditioning necessary
  • Samples are representative
  • Ready to use filtration membrane after hydrophilization
  • Robust filtration membrane which can be reused after cleaning
  • Installation in 12 and 19 mm top ports as well as 19 and 25 mm sideports
  • Sampling system for the on-line analyzer ProcessTRACE, MultiTRACE and ProcessMonitor
  • Process control via on-line HPLC
  • Sampling system for other analyzer
  • Wihtdrawal of off-line samples in connection with an external pump

Technical data:

  • Sterile in-line filtration technique
  • Prefabricated tubular polypropylene membrane
  • Autoclavable in situ at 125 °C and 1.5 bar
  • Membrane sealing with PTFE ferrules
  • Flow rate of 1 to 1.5 ml /min
  • 3 ml volume sampling per measurement in continuous process
  • Length of the extension module 100 mm; a maximum of 4 extension modules is recommended

News

08.07.10

New phone and telefax number

Phone no.: +49.531.209008-0 

TRACE Analytics has a new phone and telefax number...

11.03.10

Events 2010

Jahrestagung der Biotechnologen

Cooperation Forum "Biopharmaceuticals"

01.06.10

Publications

On-line Glucose Measurement in Disposable Bioreactors

08.05.09Certification
Certification according to ISO 9001:2008